beautypg.com

5 troubleshooting – Eppendorf FemtoJet User Manual

Page 36

background image

34

5 Troubleshooting

Error 10

Supply pressure cannot be reached

The compressor runs without the
maximum supply pressure being
reached.

Possible cause:
High compensation pressure when the
capillary holder is not attached or is
poorly attached.

Technical cause:
Compressor does not start up. Leak in
compressor or pressure accumulator.

Solution:
Check the capillary holder and tube for
leaks or tighten up the tube connections.
Service.

Error 11

Supply pressure is too high

Supply pressure has exceeded the
maximum permissible supply pressure.

Possible cause:
supply pressure is not relieved by the
relevant standby function.
Technical defect.

Solution:
Switch off using the

key (hold

down for approximately two seconds).
Service.

Error 12

Supply pressure does not drop far
enough

Possible cause: Technical defect

Solution: Service

ERROR 010

Supply pressure

cannot be reached

INJECT = Continue

ERROR 011

Supply pressure is

too high

INJECT = Continue

ERROR 012

Supply pressure is

does not fall

INJECT = Continue

05_Fehler_en.fm Seite 34 Mittwoch, 27. September 2006 5:14 17