5 Troubleshooting
Error 10
Supply pressure cannot be reached
The compressor runs without the
maximum supply pressure being
reached.
Possible cause:
High compensation pressure when the
capillary holder is not attached or is
poorly attached.
Technical cause:
Compressor does not start up. Leak in
compressor or pressure accumulator.
Solution:
Check the capillary holder and tube for
leaks or tighten up the tube connections.
Service.
Error 11
Supply pressure is too high
Supply pressure has exceeded the
maximum permissible supply pressure.
Possible cause:
supply pressure is not relieved by the
relevant standby function.
Technical defect.
Solution:
Switch off using the
key (hold
down for approximately two seconds).
Service.
Error 12
Supply pressure does not drop far
enough
Possible cause: Technical defect
Solution: Service