3 Mode of operation
Auto / Manu
To switch between manual and
automatic, time-controlled injection.
In the case of manual injection, time
is counted upwards from 0.0 seconds.
The parameter t
i
is not displayed in the
bottom two lines in order to emphasize
that the time cannot be set via the
variable regulators in this case.
Count
To reset to zero the counter for injections
which have been actuated.
Menu
To switch the device to menu control;
selected functions may be executed.
In the first line, FUNC indicates that the
device is in the function menu and shows
which function number is active.
The second line contains the function
description (e.g. "Select pressure unit").
The third line requests an action (e.g.
press the
key to change the
pressure unit). Current status: [hPa]
The two bottom lines show information
about which action is assigned to which
variable regulator (e.g. other functions
can be selected using left-hand variable
regulator p
i
).
MANU - INJECT
n= 0
0.4 s
pi[hPa]
pc[hPa]
150
50
FUNC 1
Select pressure unit
INJECT = PSI
[hPa]
Turn pi to
select function