5 troubleshooting – Eppendorf FemtoJet User Manual
Page 40
38
5 Troubleshooting
Error 28
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
Error 30
Sensor offset supply pressure is too low
Possible cause: Technical defect
Solution: Service
Error 31
Sensor offset supply pressure is too high
Possible cause: Technical defect
Solution: Service
Error 32
Sensor offset working pressure is too low
Possible cause: Technical defect
Solution: Service
Error 33
Sensor offset working pressure is too
high
Possible cause: Technical defect
Solution: Service
ERROR 028
Working pressure
leak is too high
INJECT = Continue
ERROR 030
Sensoroffset supply
is too low
INJECT = Continue
ERROR 031
Sensoroffset supply
is too high
INJECT = Continue
ERROR 032
Sensoroffset work
is too low
INJECT = Continue
ERROR 033
Sensoroffset work
is too high
INJECT = Continue
05_Fehler_en.fm Seite 38 Mittwoch, 27. September 2006 5:14 17