5 Troubleshooting

Error 28

Working pressure leak rate is too high

Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.

Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.

Error 30

Sensor offset supply pressure is too low

Possible cause: Technical defect

Solution: Service

Error 31

Sensor offset supply pressure is too high

Possible cause: Technical defect

Solution: Service

Error 32

Sensor offset working pressure is too low

Possible cause: Technical defect

Solution: Service

Error 33

Sensor offset working pressure is too
high

Possible cause: Technical defect

Solution: Service