5 Troubleshooting

5.2 Warnings

As a rule, warnings are emitted if certain operating conditions are not (yet)
reached or if an action is not possible.

Acknowledge the warning with the

key and the unit will return to

normal status.

Warning 1

Supply pressure has not yet been
reached.

The minimum supply pressure required
for an injection or the Clean function was
not present when the function was
triggered.

Possible cause:
Multiple triggering of Clean or very rapid
injection at a high injection pressure.

Solution:
Wait until the supply pressure has built up
and trigger again.

Warning 2

Compensation pressure cannot be
maintained.

Possible cause:
Capillary or capillary holder not attached
or only connected loosely.

Solution:
Check all connections and tighten.
Activate Function 0 to change the
capillary and when the capillary holder is
open.

WARNING 001

Please wait for

supply pressure

INJECT = Continue

WARNING 002

Check capillary

e.g. broken/missing ?

INJECT = Continue