Appendix A

If the command is faulty, an error message is returned instead of the
answer:

"ERR=eee"
ERR:

Error

=: Separation
eee:

Error number

Lower and upper case letters can be used.

Command to enter
parameters

[C]

Code

[xxx]

Input
parameter
[ssssssssss]

Meaning of
input parameters

Short acoustic signal

003

1 – 254

Number of signals

Long acoustic signal

004

1 – 254

Number of signals

Automatic injection
(inject at t

i

)

010

0; 1 – 999

Injection time
in 1/10 seconds
0: injection time set on
the FemtoJet

Set injection counter n
to 0

012

– – –

No parameter input

Perform Clean

013

1 – 100

Time in 1/10 seconds

Execute Standby

014

0 or 1

0 = short
1 = long

Switch acoustic signal
on/off

040

0 or 1

Status:
0 = off / 1 = on

Set pressure unit

041

0 or 1

Unit:
0 = hPa / 1 = psi

Set injection time t

i

042

1 – 999

Injection time
in 1/10 seconds

A_Bedien_Computer_en.fm Seite 50 Mittwoch, 27. September 2006 5:26 17