5 Troubleshooting
Error 21
Injection pressure has not been reached
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
Error 22
Compensation pressure has not been
reached
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
Error 23
Increase in working pressure is smaller
than required
Possible cause: Technical defect
Solution: Service
Error 24
No increase in working pressure
Possible cause: Technical defect
Solution: Service
ERROR 021
Inject pressure
cannot be reached
INJECT = Continue
ERROR 022
Compensation pressure
cannot be reached
INJECT = Continue
ERROR 023
Inject n pressure
overflow too large
INJECT = Continue
ERROR 024
Working pressure
does not rise
INJECT = Continue