beautypg.com

3 mode of operation – Eppendorf FemtoJet User Manual

Page 18

background image

16

3 Mode of operation

Standby

in the case of a brief interruption

When key is pressed briefly:
The device switches to the Standby
mode

without

deairing. The illumination

of the display switches off. "STANDBY"
appears in the display with the suffix
"Supply pressure = rxyz" as a code for
system pressure p

s

being maintained.

Application:
For temporary interruptions or use of

as a Reset function.

When key is pressed again:
The device is reactivated and runs
through a short warm-up routine.
This lasts approximately one minute.

Standby

with deairing of pressure accumulator

Hold down the key (approx. 2 seconds;
the display moves to "STANDBY"):

The device switches to the Standby
mode. The pressure accumulator is
completely deaired, the illumination of
the display switches off. "STANDBY"
appears in the display with the note
"Supply pressure = x" (x ~ 0).

A noise can be heard at this stage.

Application:
For correct shutdown, with deairing of the
entire pressure system in the event of
longer periods of non-use or standstill
(weekend or storage) with subsequent
switch-off using the mains switch.

Press the key again:
The device is reactivated and runs
through the warm-up routine.
This takes approximately three minutes.

STANDBY

Supply pressure

6789

Hold standby

to exhaust

STANDBY

Supply pressure

0