3 Mode of operation

3.7.5 Constant pressure

3.7.6 Dewatering the pressure accumulator

FUNC 4
This function can be used to apply a
constant working pressure p

w

in the

range from 5 to 6,000 hPa (87.0 psi) at
the pressure outlet. The pressure unit psi
can also be used.

The working pressure p

w

is set with

the right-hand variable regulator (p

c

),

the specified value is shown in large
characters. The actual value is shown
in the bottom right of the display.

The Clean function can be activated. The
pressure of the Clean function is shown
in the bottom right of the display in the
form of p

w

.

FUNC 5

Briefly opens the dewatering valve
of the pressure accumulator; any
condensed humidity present is blown
away.

When this function is performed, the
noise of air being blown away can be
heard.

This function is automatically performed
every time the unit is switched on. It can
be performed manually as described if
condensation is suspected to be present.

FUNC 4
Continuous flow
INJECT = start

Turn pi to
select function

FUNC 4
Continuous flow

pw= 50 hPa

Turn pi to

pw[hPa]

select function

50

FUNC 5
Drain pressure supply
INJECT = start

Turn pi to
select function