beautypg.com

3 error messages, 5 troubleshooting – Eppendorf FemtoJet User Manual

Page 35

background image

33

5 Troubleshooting

5.3 Error messages

Error messages are emitted if certain operating conditions are not
reached or the device records a deviation from specified conditions.

Acknowledge the error message with the

key and the unit will

return to normal status. The measures described under “Solution” should
be carried out. If the error message recurs after acknowledgement with
the

key, inform Service.

Warning 3

Micromanipulator does not react

A micromanipulator is connected but
does not react to a start pulse from the
FemtoJet

.

Possible cause:
Micromanipulator is switched off.
INJECT function on manipulator is not
released.
No limit is set on the micromanipulator.
In the INJECT function on the micro-
manipulator, synchronization is set to
limit (SYNC = LIMIT).

Solution:
Switch on the micromanipulator and
make it operational.
Disconnect the cable to micro-
manipulator.
With SYNC = LIMIT, trigger semi-auto-
matic injection via the micromanipulator
only.

WARNING 003

Switch on manipulator

or disconnect cable

INJECT = Continue

05_Fehler_en.fm Seite 33 Mittwoch, 27. September 2006 5:14 17