4 information on working practices, 3 mode of operation – Eppendorf FemtoJet User Manual
Page 21
19
3 Mode of operation
Other accessories
Grip head 0,
incl. 2 sets of O-rings and sealing washer.
Using grip head 0 supplied, capillaries with an outer
diameter of 1.0 mm to 1.1 mm can be used. In the
case of outer diameters which differ from these
measurements, the relevant grip heads can be orde-
red as non-standard accessories for the universal
capillary holder (see Section 7, Ordering informa-
tion). The transparent grip head allows the user to
visually monitor the position of the capillary.
3.4 Information on working practices
The following pressures can be used:
Name
Pressure range
Comments
Compensation
pressure p
c
0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi
Injection pressure p
i
0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi
Clean pressure
maximum
up to 7,000 hPa
maximum up to 102 psi
Constant working
pressure p
w
0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi
Select Function 4 via
the
key.
System pressure p
s
[supply pressure]
maximum
up to 7,000 hPa
maximum up to 102 psi
This is used directly for
.
In certain phases of
initialization, the
system pressure may
temporarily increase
to a maximum of
8,000 hPa (117 psi).
03_Bedien_en.fm Seite 19 Mittwoch, 27. September 2006 5:12 17