3 Mode of operation

Other accessories

Grip head 0,
incl. 2 sets of O-rings and sealing washer.

Using grip head 0 supplied, capillaries with an outer
diameter of 1.0 mm to 1.1 mm can be used. In the
case of outer diameters which differ from these
measurements, the relevant grip heads can be orde-
red as non-standard accessories for the universal
capillary holder (see Section 7, Ordering informa-
tion). The transparent grip head allows the user to
visually monitor the position of the capillary.

3.4 Information on working practices

The following pressures can be used:

Name

Pressure range

Comments

Compensation
pressure p

c

0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi

Injection pressure p

i

0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi

Clean pressure

maximum
up to 7,000 hPa
maximum up to 102 psi

Constant working
pressure p

w

0; 5 – 6,000 hPa
0.00; 0.07 – 87.0 psi

Select Function 4 via
the

key.

System pressure p

s

[supply pressure]

maximum
up to 7,000 hPa
maximum up to 102 psi

This is used directly for

.

In certain phases of
initialization, the
system pressure may
temporarily increase
to a maximum of
8,000 hPa (117 psi).

03_Bedien_en.fm Seite 19 Mittwoch, 27. September 2006 5:12 17