3 Mode of operation

Slight deviations between the nominal value (specified value) and
displayed compensation pressure p

c

(actual value) may occur for

technical reasons. The relevant value is the actual value measured.

3.1.2 Variable regulators

The variable regulators are set so that when they are turned slowly, the
values are changed by 1 hPa / 0.1 second per notch (0.01 – 0.02 psi).
If the regulators are turned faster, the values are altered in larger
increments.

The following settings are used as a
basis for initial injection experiments
using Femtotip

/ Femtotip

II:

Display:

p

i

left-hand variable regulator: 150 hPa

t

i

central variable regulator:

0.5 s

p

c

right-hand variable regulator: 50 hPa

Variable regulators:

AUTO

n= 0

pc= 50 hPa

pi[hPa]

ti[s]

pc[hPa]

150

0.5

50