5 Troubleshooting
5.1 Potential errors
–
Pressure loss due to capillary not fitted with capillary holder
connected.
Solution: Connect the capillary or select Function 0, Change capillary,
via the
key and variable regulator p
i
.
–
Capillary is blocked.
Solution: Press the
key for long enough.
Change capillary.
Purify injection fluid (e.g. by means of centrifugation).
–
Compressor runs for a long time.
Solution: Take a short break.
Do not use the
key too often within a short period.
If possible, set a lower compensation pressure p
c.
Perform self-test (Section 5.1.1).
–
Valve rattles
Solution: Connect tube to filled capillary. Pressure control is designed
for the tube connected.
–
No communication with the micromanipulator.
Solution: Check cables and connections.
5.1.1 Self-test
FUNC 6
The FemtoJet
is checked for leaks in
the self-test.
This self-test can also be used if a leak is
suspected.
The test runs automatically.
The injection tube must be connected to
a closed capillary (e.g. melted closed or
filled) for the pressure control to work
properly.
FUNC 6
Self-Test
INJECT = Complete
CLEAN = Leak only
Turn pi to
select function