beautypg.com

5 troubleshooting – Eppendorf FemtoJet User Manual

Page 33

background image

31

5 Troubleshooting

key: Complete self-test

Checks compressor run time up to
the point when a specific pressure
is reached and thus for leaks in the
area of the compressor.

In the case of two different
pressures, checks for leaks in the
pressure accumulator (p

s

) and for

the internal tightness of the valves.

Checks for leaks in the area of
working pressure (p

w

) in the case

of two different pressures.

Checks behavior during injection
at different pressures (short
continuous test).

If certain values are outside the specified
range, an error message is emitted. If this
recurs, inform Service.

key: Self-test to search for leaks

Checks for leaks in the pressure
accumulator (p

s

) in the case of two

different pressures, as well as for
inner tightness of the valves.

Checks for leaks in the area of wor-
king pressure (p

w

) in the case of two

different pressures.

If certain values are outside the specified
range, an error message is emitted. If this
recurs, inform Service.

FUNC 6
Self-Test
Compressor

_____ s

Ps leak

_____

_____ hPa

Pw leak

_____

_____ hPa

Injection _____

_____

hPa

ps=7000

pw=

18 hPa

FUNC 6
Self-Test

Ps leak

_____

_____ hPa

Pw leak

_____

_____ hPa

ps=7000

pw=

18 hPa

05_Fehler_en.fm Seite 31 Mittwoch, 27. September 2006 5:14 17