3 Mode of operation

3.7.1 Changing the capillaries

3.7.2 Setting the pressure unit

3.7.3 Switching on/off the loudspeaker

3.7.4 Switching on/off the background illumination of the display

FUNC 0
When this function is called up, the
injection tube is deaired and any
pressure present in the system is let out.
Once the capillary has been changed,

exit the function using the

key or

variable regulator p

i

; pressure control will

restart.

FUNC 1
The other pressure unit is selected using

the

key and all pressure values

are converted. The pressure unit
currently selected (in this case: hPa)
is underlined.

FUNC 2
Use the

key to switch the

loudspeaker on or off. The currently
selected setting (in this case: ON) is
underlined. Acoustic error messages are
not emitted if the setting
is OFF.

FUNC 3
Use the

key to switch the

illumination on or off. The currently
selected setting (in this case: ON) is
underlined.

In the normal operating mode,
illumination switches off after three
minutes if no operating control has been
actuated. When an operating control is
actuated, the illumination lights up again.

FUNC 0
Change capillary

Capillary may be
changed now

Turn pi to
select function

FUNC 1
Push Inject to select
Pressure Unit

hPa PSI

Turn pi to
select function

FUNC 2
Push Inject to select
Beeper ON OFF

Turn pi to
select function

FUNC 3
Push Inject to select
Illumination

ON OFF

Turn pi to
select function

03_Bedien_en.fm Seite 27 Mittwoch, 27. September 2006 5:12 17