beautypg.com

5 troubleshooting – Eppendorf FemtoJet User Manual

Page 39

background image

37

5 Troubleshooting

Error 25

No drop in working pressure

Possible cause: Technical defect

Solution: Service

Error 26

Working pressure leak rate is too high

Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.

Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.

Error 27

Working pressure leak rate is too high

Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.

Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.

ERROR 025

Working pressure

does not fall

INJECT = Continue

ERROR 026

Working pressure

leak is too high

INJECT = Continue

ERROR 027

Working pressure

leak is too high

INJECT = Continue

05_Fehler_en.fm Seite 37 Mittwoch, 27. September 2006 5:14 17