5 troubleshooting – Eppendorf FemtoJet User Manual
Page 39

37
5 Troubleshooting
Error 25
No drop in working pressure
Possible cause: Technical defect
Solution: Service
Error 26
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
Error 27
Working pressure leak rate is too high
Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.
Solution:
Check capillary. Check capillary holder
and tube for leaks or tighten up tube
connections.
Service.
ERROR 025
Working pressure
does not fall
INJECT = Continue
ERROR 026
Working pressure
leak is too high
INJECT = Continue
ERROR 027
Working pressure
leak is too high
INJECT = Continue
05_Fehler_en.fm Seite 37 Mittwoch, 27. September 2006 5:14 17
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (64 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (142 pages)
- Mastercycler nexus (118 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (72 pages)
- AF2200 Plate Reader (78 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (25 pages)
- MiniSpin (20 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (80 pages)
- 5418 Centrifuge (48 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (95 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)