5 Troubleshooting

Error 13

Supply pressure does not rise

Possible cause: Technical defect

Solution: Service

Error 14

Massive leak in pressure accumulator

Possible cause: Technical defect

Solution: Service

Error 15

Leak in pressure accumulator

Possible cause: Technical defect

Solution: Service

Error 16

Leak through Valve V1 is too large

Possible cause: Technical defect

Solution: Service

Error 20

Working pressure has not been reached

Compressor runs without supply
pressure being reached within a specific
time.

Possible cause:
Capillary holder not attached or poorly
attached. (Broken) capillary with large
opening.
Technical defect.

Solution:
Check the capillary. Check the capillary
holder and tube for leaks or tighten up the
tube connections.
Service.

ERROR 013

Supply pressure is

does not raise

INJECT = Continue

ERROR 014

Supply pressure

leak is too high

INJECT = Continue

ERROR 015

Supply pressure

leak is too high

INJECT = Continue

ERROR 016

Valve 1 pressure

leak is too high

INJECT = Continue

ERROR 020

Working pressure

cannot be reached

INJECT = Continue