beautypg.com

WIKA CPG2500 User Manual

Page 72

background image

72

Mensor/WIKA Operating Instructions Series CPG 2500

Digital Pressure Gauge
SERIES CPG 2500

9.2 Spare Parts

Table 9.2 lists the spare parts for the CPG 2500 that can be ordered from Mensor.

Table 9.2 Spare Parts List

Part Description

Part Number

Manual

0017851001

Fuse

Littelfuse® 0454 002

Power Supply

4901000029

Fitting adapter 7/16-20 to 1/8 FNPT

6000602015

Fitting adapter 7/16-20 to 1/4 FNPT

0012688001

O-ring seals for 7/16-20 fitting

4250010020

Kit - Rack Mount adapter, single unit

0017839001

9.3 Replacing the Fuse

A spare fuse has been provided and is located on the circuit board as shown below.

!

Caution

In the event the fuse requires replacement, power down the instru-
ment before proceeding!

1. Power off the instrument.
2.

Remove the instrument
cover and locate the circuit
board.

3.

Remove the spare fuse.
Replace the fuse of F1 with
the spare.

4.

If the instrument blows the
replacement fuse, please
return the instrument for
service.