40

Mensor/WIKA Operating Instructions Series CPG 2500

Digital Pressure Gauge
SERIES CPG 2500

A?

n.nnnnne+nn

Returns the A channel pressure reading.

AR?

n.nnnnne+nn

Returns the A channel rate.

BARO?

n.nnnnne+nn

Returns reading from barometric sensor

B?

n.nnnnne+nn

Returns the B channel pressure reading.

BR?

n.nnnnne+nn

Returns the B channel rate.

CALDISABLE

yes,no

Sets whether or not calibration of the ac-
tive channel is disabled.

CALDISABLE?

(yes or no)

Returns whether or not calibration of the
active channel is disabled.

CERR

none

Clears the error queue.

CHAN

A or B

Sets the active channel on the instrument.

CHAN?

A or B or D or
Baro

Returns which channel is active.

CHAN D

none

Sets the active channel to the Delta chan-
nel.

CHANFUNC

Press, peak, rate, baro

Sets the alternate function mode of the
active channel.

CHANFUNC?

CCCCC…

Returns the alternate function mode of the
channel.

DECPT?

n

Returns the number of decimal points for
the active channel (see RESOLUTION).

DEFAULT

none

Sets the default values.

DELTAFUNC A-B

none

Sets the delta to be the result of the chan-
nel A reading - channel B reading.
(Chan D must be in the active channel).

DELTAFUNC B-A

none

Sets the delta to be the result of the chan-
nel B reading - channel A reading.
(Chan D must be in the active channel).

DELTAFUNC A+B

none

Sets the delta to be the result of the chan-
nel A reading + channel B reading.
(Chan D must be in the active channel).

DELTAFUNC?

{A-B or B-A}

Returns delta function as one of the above
options. (Chan D must be in the active
channel).