Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual
Page 43
Mensor/WIKA Operating Instructions Series CPG 2500
43
Digital Pressure Gauge
SERIES CPG 2500
RWINDOW?
Returns rate exponential filter window.
SBAUD
9600, 19200, 38400, 57600
Sets the serial baud rate.
SBAUD?
Returns the serial baud data.
SCREENSRC
A, B, or A,B, or B,A, or
Delta,A, or Delta,B
Sets the CPG 2500 display mode.
SCREENSRC?
Delta,B}
Returns screen source as one of the
above options.
SDATA
7 or 8
Sets the serial data bits.
SDATA?
Returns the serial data bits number.
SENSORID?
XXXXXX,VER V.VV
Returns the active sensor’s serial number
and firmware version.
SPAN
desired pressure or ?
Sets span on active channel or for ?,
clears previous value, must be >50% FS
and has a 1% limit.
SPAN?
Returns span scale factor for active chan-
nel.
SPARITY
even, odd, none
Sets the serial parity.
SPARITY?
Returns the serial parity.
SSTOP
1 or 2
Sets the serial stop bits.
SSTOP?
Returns the serial stop bits.
UNITS
units code or text in Table
7.1.3.6.
Sets the engineering units for the active
channel.
UNITS?
Returns the pressure units of the active
channel in a text string.
UNLOCK
none
Releases Acquire locks. See ACQUIRE?
and RELEASE?
WINDOW
value in current units
Sets the exponential filter window for the
active channel.
WINDOW?
Returns the exponential filter window for
the active channel.
ZERO
desired pressure or ?
Sets zero to set pressure or for ?, clears
previous value.
ZERO?
Returns zero offset for active channel.