Mensor/WIKA Operating Instructions Series CPG 2500

55

Digital Pressure Gauge

SERIES CPG 2500

The ‘Q’ Command Data/Output Format

n

Description

Output Format

0

Pressure reading in the se-
lected units. This is the default
output format.

NNNNNNNwhere each N is a number (0
through 9), +,-,decimal point, or a space. If the pres-
sure rate or peak monitor option is being used the
output format is NNNNNNN,NNNNNNN. The
data before the comma is the pressure and the data
after the comma is the pressure rate or peak.

1

Raw A/D readings (for factory
use)

nnnnnnn,nnnnnnn where each n is a number
(0 through 9) or a space.

2

Unit ID

Unit ID for the CPG 2500

3

Pressure range and measure-
ment units

NNNNNNN,NNNNNNN, @@@@@@@ where
each N is a number (0 through 9), decimal point, or a
space and each @ is an alpha-numeric character. The
output string represents the minimum pressure, maxi-
mum pressure and the pressure units.

4

Error status code

CPG 2500’s error codes

5

Calibration data

not emulated

6

Pressure rate or peak indica-
tion

nnnn where each n is a number (0 through 9) valid only
if optional pressure rate or peak is enabled. The mean-
ing of each digit is shown under the related feature in
the Options section of the DPG II manual.

7

Zero and span corrections

ZZZZZZZ,SSSSSSS where ZZZZZZZ is the
zero correction and SSSSSSS is the span correction.

8

Calibration coefficients

not emulated

S Command: same as SPAN command.

T Command: used with RATE option; similar to DISPLAY command.