Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual
Page 55
Mensor/WIKA Operating Instructions Series CPG 2500
55
Digital Pressure Gauge
SERIES CPG 2500
The ‘Q’ Command Data/Output Format
n
Description
Output Format
0
Pressure reading in the se-
lected units. This is the default
output format.
NNNNNNN
through 9), +,-,decimal point, or a space. If the pres-
sure rate or peak monitor option is being used the
output format is NNNNNNN,NNNNNNN
data before the comma is the pressure and the data
after the comma is the pressure rate or peak.
1
Raw A/D readings (for factory
use)
nnnnnnn,nnnnnnn
(0 through 9) or a space.
2
Unit ID
Unit ID for the CPG 2500
3
Pressure range and measure-
ment units
NNNNNNN,NNNNNNN, @@@@@@@
each N is a number (0 through 9), decimal point, or a
space and each @ is an alpha-numeric character. The
output string represents the minimum pressure, maxi-
mum pressure and the pressure units.
4
Error status code
CPG 2500’s error codes
5
Calibration data
not emulated
6
Pressure rate or peak indica-
tion
nnnn where each n is a number (0 through 9) valid only
if optional pressure rate or peak is enabled. The mean-
ing of each digit is shown under the related feature in
the Options section of the DPG II manual.
7
Zero and span corrections
ZZZZZZZ,SSSSSSS
zero correction and SSSSSSS is the span correction.
8
Calibration coefficients
not emulated
S Command: same as SPAN command.
T Command: used with RATE option; similar to DISPLAY command.