beautypg.com

Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual

Page 58

background image

58

Mensor/WIKA Operating Instructions Series CPG 2500

Digital Pressure Gauge
SERIES CPG 2500

7.1.3.8.4 General SCPI WIKA Commands

STATus

:OPERation

:CONDition?

Returns an integer value representing instru-
ment status that can be decoded.
Bit 0: Zeroing active.
Bit 1: Control Setpoint has not been reached.
Bit 2: Reserved 0.
Bit 3: Reserved 0.
Bit 4: Measuring. The instrument is actively
measuring.

MEASure

[:PRESsure]

[z]?

Returns the pressure in the current units from
the specified sensor, or active sensor if [z] is
omitted.

:TEMPerature[z]?

Returns the temperature in deg. C. from the
specified sensor, or active sensor if [z] is omit-
ted.

:RATE[z]?

Returns the pressure rate in the current units
from the specified sensor, or active sensor if [z]
is omitted.

:BAROmetric?

Returns the barometric pressure in the current
unit, if a barometer sensor is installed.

CALibration

:MODE?

Returns 1 if the sensor has been calibrated or 0
if not.

:DATE?

Returns the stored calibration date.

:DATE

Sets the calibration date (YYYY,MM,DD).

:ZERO?

Returns current zero point correction in bar.

:ZERO

Sets the zero so the currently measured

pressure

matches

the

commanded

value

in

the

current

unit.

Limited

to

+/-5%

FS.

ZERO:RUN

Vents the instrument and sets zero so the

currently

measured

pressure

=

0.