Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual
Page 51
Mensor/WIKA Operating Instructions Series CPG 2500
51
Digital Pressure Gauge
SERIES CPG 2500
SPAN
SPAN sp
Use the SPAN command to correct the pressure
reading at FS. This command first requires the
PW password to be issued. Send the true pres-
sure value while maintaining at least 50% of the
FS pressure on the pressure port. (Refer to the
Maintenance and Calibration sections for details
on calibrating an instrument.)
Where
engineering units.
SPAN?
SPAN? lf
The SPAN query will return the span correction
scale factor. Returns
1.10000.
TARE
TARE sp
TARE sets the tare offset in the current engineering
units. Either the PWT or the PW password must
be sent before the first occurrence of the TARE
command. Where
7.0000 psi.
TARE?
TARE?lf
Returns the current tare calibration variable.
Returns
current TARE offset in the current units.
TYPE?
type?lf
Returns the type of pressure sensor in the instru-
ment:
ABSOLUTE PRESSURE
sensor,
or: GAUGE PRESSURE
sure sensor.
UNITS
UNITS sp
This command selects the engineering units to be
output on the bus and the display for all subse-
quent pressure readings.
Where
number from the ‘n’ column of Table 7.1.3.6.
UNITS?
units?lf
This query returns the units code and the ASCII
string for the units as
where
(see Table 7.1.3.6) and where
sponding ascii string listed under ‘Output Format.