Digital pressure gauge series cpg 2500 – WIKA CPG2500 User Manual
Page 66
66
Mensor/WIKA Operating Instructions Series CPG 2500
Digital Pressure Gauge
SERIES CPG 2500
9000
2743.20
10.50483
21.38811
543.25820
10000
3048.00
10.10644
20.57696
522.65508
15000
4572.00
8.29351
16.88579
428.89932
20000
6096.00
6.75340
13.75009
349.25242
25000
7620.00
5.45352
11.10351
282.02934
30000
9144.00
4.36410
8.88542
225.68990
*Standard sea level pressure: 0 feet = 29.92126 inches HgA (0 meters = 1013.25 mbar).
8.2.2 Calibration
The Barometer can be calibrated in exactly the same manner as the other installed sen-
sors as described in Section 10, Calibration.
8.2.3 Specifications
Accuracy: 0.01% of reading. Uncertainties include all pressure effects, temperature
effects over the calibrated range and calibration stability for six months after
re-zeroing.
Pressure Range: The Barometer is calibrated from 11 to 17 psia.
Resolution: 6 digits.
- 890.09.2190 (44 pages)
- A-10 (96 pages)
- A2G-50 (52 pages)
- A2G-55 (36 pages)
- AC-1 (88 pages)
- C-2 (24 pages)
- D-10-7 (112 pages)
- D-20-9 (51 pages)
- D-20-9 (35 pages)
- DG-10 (112 pages)
- DP-10 (44 pages)
- DPT-10 (96 pages)
- DPT-10 (92 pages)
- GCS-1 (76 pages)
- HP-2 (84 pages)
- IL-10 (31 pages)
- IPT-10 (56 pages)
- IPT-10 (48 pages)
- IPT-10 (28 pages)
- IPT-10 (36 pages)
- IS-20-F (7 pages)
- IS-20-F (43 pages)
- IS-20-H (7 pages)
- UT-10 (42 pages)
- UT-10 (52 pages)
- IUT-10 (78 pages)
- LH-10 (60 pages)
- LH-20 (60 pages)
- LS-10 (60 pages)
- MG-1 (92 pages)
- MH-1 (11 pages)
- MH-2 (9 pages)
- MHC-1 (84 pages)
- N-10 (35 pages)
- O-10 (108 pages)
- OT-1 (60 pages)
- P-30 (92 pages)
- PSA-31 (124 pages)
- PSD-30 (128 pages)
- R-1 (92 pages)
- S-10 (35 pages)
- S-11 (102 pages)
- S-20 (52 pages)
- S-20 (96 pages)
- SL-1 (21 pages)