9000

2743.20

10.50483

21.38811

543.25820

10000

3048.00

10.10644

20.57696

522.65508

15000

4572.00

8.29351

16.88579

428.89932

20000

6096.00

6.75340

13.75009

349.25242

25000

7620.00

5.45352

11.10351

282.02934

30000

9144.00

4.36410

8.88542

225.68990

*Standard sea level pressure: 0 feet = 29.92126 inches HgA (0 meters = 1013.25 mbar).

8.2.2 Calibration

The Barometer can be calibrated in exactly the same manner as the other installed sen-
sors as described in Section 10, Calibration.

8.2.3 Specifications

Accuracy: 0.01% of reading. Uncertainties include all pressure effects, temperature

effects over the calibrated range and calibration stability for six months after
re-zeroing.

Pressure Range: The Barometer is calibrated from 11 to 17 psia.

Resolution: 6 digits.