Using the agilent e1439 – Agilent Technologies VXI E1439 User Manual
Page 62
52
Using the Agilent E1439
Fiber Optic Interface
The following shows two E1439D modules in an append chain transmitting data to a fiber
receiver when flow control is on.
Fiber TX
Fiber RX
VME BUS
LBUS
ADC
DIGITAL
FILTERS
FIFO
Fiber TX
Fiber RX
VME BUS
LBUS
ADC
DIGITAL
FILTERS
FIFO
Fiber
TX
Fiber
RX
Processing
DATA
E1438D / E1439D
Fiber Receiver
E1438D / E1439D
1 KB
FIFO
1 KB
FIFO
1 KB
FIFO
Fiber Interface Setup
First E1439D in chain
Next E1439D in chain
Mode
Generate
Mode
Append
Rate
106 or 250 MBs
Rate
same as module to left
BOF
ON
BOF
ON
1
1. The final module in an append chain may require BOF to be Off for compat-
ibility with data receivers that cannot process BOFs.
CRC
ON
CRC
ON
Flow Control
Copy
Flow Control
No Copy
2
2. Set intermediate modules to Copy and the last module to No Copy.
Epoch Generate
ON
Epoch Generate
ON
Epoch Size
Divisible by 4; must match blocksize
Epoch Size