48
Using the Agilent E1439
Fiber Optic Interface
Generate
If flow control is off, Generate fiber mode transmits filtered ADC data over the fiber interface as
soon as data is available. ADC data is not available via any other data port and received optical
data is ignored. The following illustration shows an E1439D transmitting data when flow control
is turned off.
Fiber Interface Setup
Fiber Mode
Generate
Rate
106 or 250 MBs
BOF
Optional
CRC
ON
1
1. Some receivers may
require CRC to be off
for compatibility
Flow Control
OFF
Epoch Generate
Optional
Epoch Size
Divisible by 4
Fiber TX
Fiber RX
VME BUS
LBUS
ADC
DIGITAL
FILTERS
FIFO
E1438D / E1439D
1 KB
FIFO
This manual is related to the following products: