beautypg.com

Age1439_fiber_signal_get – Agilent Technologies VXI E1439 User Manual

Page 125

background image

115

Agilent E1439 Programmer's Reference

Functions listed alphabetically

age1439_fiber_signal_get

Returns a value indicating whether or not an optical signal is detected by the optical fiber
interface receiver.

VXIplug&play Syntax

#include "age1439".h

ViStatus age1439_fiber_signal_get(ViSession id, ViPInt16 fiberSignalPtr);

Parameters

id

is the VXI instrument session pointer returned by the age1439_init function.

fiberSignalPtr

returns a value indicating whether or not an optical signal has been detected by the fiber interface
receiver.

AGE1439_NO_FIBER_SIGNAL

indicates no optical signal has been detected by the fiber

interface receiver.

AGE1439_FIBER_SIGNAL_PRESENT

indicates an optical signal has been detected by the fiber

interface receiver.

Return Value

AGE1439_SUCCESS indicates that a function was successful.

Values other than AGE1439_SUCCESS indicate an error condition or other important status
condition. To determine the error message, pass the return value to

“age1439_error_message” on

page 102

.

See Also

“age1439_init” on page 132

,

“age1439_fiber_LED_get” on page 110

,

“age1439_status_get” on

page 176

This manual is related to the following products: