Campbell Scientific CPEC200 Closed-Path Eddy-Covariance System User Manual

Page 73

background image

Appendix B. Public Variables

TABLE B-1. CPEC200 public variables

When Defined

Usage

Variable Name

Units

Description

VALVE_MODULE

INFO/USER valve_number

Valve number (0 to 6);

determines which zero/span

valve is selected

VALVE_MODULE

USER

DO_P_offset

Set = True to measure the

pressure offset

VALVE_MODULE

INFO

press_offset

kPa

Sample cell pressure measured

with no flow; used to infer

valve flow

VALVE_MODULE

INFO/USER valveTctl_ON

Set = True to enable valve

module temperature control

VALVE_MODULE

INFO

valve_tmpr

ºC

Temperature of the valve

module

VALVE_MODULE

DIAG

valve_tmprOK

Boolean flag: True if the valve

temperature is within its

operating range

VALVE_MODULE

INFO

valveHeat_ON

Boolean flag: True if the valve

heater is on

VALVE_MODULE

INFO

valveFan_ON

Boolean flag: True if the valve

fan is on

VALVE_MODULE

INFO

valve_flow

LPM

Zero/span flow

VALVE_MODULE

DIAG

valve_flow_OK

Boolean flag: True if

valve_flow is within the normal

range

VALVE_MODULE

INFO

valveControl

Zero/span flow control valve;

can be from 0 (closed) to 1

(open)

SCRUB_MODULE

INFO

scrub_tmpr

ºC

Temperature of the scrub

module

SCRUB_MODULE

DIAG

scrub_tmprOK

Boolean flag: True if the scrub

module temperature is within

its operating range

SCRUB_MODULE

INFO

scrubHeat_ON

Boolean flag: True if the scrub

module heater is on

SCRUB_MODULE

INFO

scrubFan_ON

Boolean flag: True if the scrub

module fan is on

SCRUB_MODULE

INFO

scrub_press

kPa

Pressure measured at the scrub

pump outlet

VALVE_MODULE

INFO/USER DO_zero

Set = True to send the CO

2

and

H

2

O zero command to the

IRGA

VALVE_MODULE

INFO/USER DO_CO2_span

Set = True to send the CO

2

span command to the IRGA

VALVE_MODULE

USER

H2OSpanDewPt

ºC

Setpoint temperature on

dewpoint generator; used for

H

2

O span

VALVE_MODULE

USER

DO_H2O_span

Set = True to send the H

2

O

span command to the IRGA

always

INFO

BattVolt

V

Supply voltage; measured

inside datalogger after diode

voltage drop

B-3