beautypg.com

3 specifications, Installation, Specifications – Campbell Scientific CPEC200 Closed-Path Eddy-Covariance System User Manual

Page 27

background image

CPEC200 Closed-Path Eddy-Covariance System

The outlet of the pump connects the

Exhaust fitting on the bottom of the pump

module enclosure. This fitting has a screen to prevent insects or debris from

entering when the pump is off.

4.3 Specifications

System

Operating temperature: –30° to +50°C

Input voltage:

10.5 to 16.0 Vdc

Power:

12 W (typical), 35 W (max, at cold startup)

System enclosure

Dimensions:

52.1 x 44.5 x 29.7 cm (20.5 x 17.5 x 11.7 in)

Weight basic system:

11.6 kg (25.5 lb)

CR3000:

1.6 kg (3.7 lb)

CFM100/NL115:

0.2 kg (0.4 lb)

Three-valve module:

1.5 kg (3.3 lb)

Six-valve module:

1.9 kg (4.2 lb)

Pump module

Cable length:

3.0 m (10 ft)

Inlet connection:

3/8-in Swagelok

®

Pressure sensor range:

15 to 115 kPa

Pumping speed:

3 to 9 LPM (automatically controlled at the

set-point, typically 7 LPM)

Dimensions:

35.6 x 29.2 x 13.5 cm (14.0 x 11.5 x 5.3 in)

Weight w/ out mounting: 5.4 kg (11.8 lb)

Valve module

Flow rate:

1 to 5 LPM (automatically controlled at the

set-point, typically 1 LPM)

Inlets

Three-valve module: Zero, CO

2

span, and H

2

O span

Six-valve module:

Zero, CO

2

spans 1 through 4, and H

2

O span

Outlets:

Analyzer and H

2

O bypass

Connections:

1/4-in Swagelok

®

Dimensions:

14.0 x 12.7 x 14.0 cm (5.5 x 5.0 x 5.5 in.)

Weight

Three-valve module: 1.5 kg (3.3 lb)

Six-valve module:

1.9 kg (4.2 lb)

EC155 and CSAT3A Specifications: see the user manual: EC155 CO

2

and

H

2

O Closed-path Gas Analyzer Manual and CSAT3 Three Dimensional Sonic

Anemometer Manual

5. Installation

The following tools are required to install the CPEC200 system in the field.

Additional tools may be required for a user-supplied tripod or tower.

9/16-in, open-end wrench

1/2-in, open-end wrench

11/16-in, open-end wrench

Adjustable wrench

17