CPEC200 Closed-Path Eddy-Covariance System

Set

valve_number to ZeroAir (1). If onsite, look at the LEDs on the valve

module to confirm the selected valve is now active. Verify

valve_status

reports Valve flow is OK. If not, check the value of

valve_flow and

troubleshoot as needed.

Verify

cpec_status reports CPEC is OK. If not, troubleshoot as needed.

There is no automatic error checking in full manual mode, so

diagnostics must be checked manually.

Watch the values of CO

2

and H

2

O. When they stabilize, set

DO_zero = True.

The IRGA will set its CO

2

and H

2

O readings to zero. Verify the values of H

2

O

and CO

2

are near zero.

Set

valve_number = None (0) to stop the flow of zero gas.

6.3.2.3 Checking and Setting the CO

2

Span

The keyboard display has menus to facilitate manual zero control for users that

are onsite. Navigate:

Manual zero/span → Manual Control → CO2 Span

Set

valve_number to one of the CO

2

span valves. If onsite, look at the LEDs

on the valve module to confirm the selected valve is now active. Verify

valve_status reports Valve flow is OK. If not, check the value of valve_flow

and troubleshoot as needed.

The 3-valve module has only one CO

2

span valve available,

CO2span1. The 6-valve module has three additional CO

2

span

valves available.

Make sure the value in

CO2_SPAN_PPM matches the CO

2

concentration of

the CO

2

span cylinder you have selected.

Verify

cpec_status reports CPEC is OK. If not, troubleshoot as needed.

There is no automatic error checking in full manual mode, so

diagnostics must be checked manually.

Watch the values of CO

2

and H

2

O. When they stabilize, set

DO_CO2_span =

True. The IRGA will set the CO

2

span to the value in

CO2_SPAN_PPM.

Verify the measured value

CO2 now matches the value of CO2_SPAN_PPM.

Set

valve_number = None (0) to stop the flow of CO

2

span gas.

6.3.2.4 Checking and Setting the H

2

O Span

H

2

O is more difficult to span than CO

2

because it is more difficult to provide a

source of air with known humidity. In most cases, a commercially available

dewpoint generator is used.

NOTE

NOTE

NOTE

38