CPEC200 Closed-Path Eddy-Covariance System

6.3.2 Full Manual Control of Zero and Span

In some cases it may be more appropriate to run the zero/span under full

manual control. This allows the user to decide how much time is required for

the zero or span gas to reach equilibrium, or to perform additional status

checking. It also allows the H

2

O to be spanned. The keyboard display has

menus to facilitate manual zero/span control for users that are onsite.

Navigate:

Manual zero/span → Manual Control

6.3.2.1 Getting Ready

There are two tasks to prepare for manual zero/span control. First, enable the

temperature control for the valves and the scrub module (if being used) as

noted above.

Second, measure the pressure offset. The keyboard display has menus to

facilitate manual zero/span control for users that are onsite. Navigate:

Manual zero/span → Manual Control → Pressure Offset

Turn the sample pump off by setting

pump_ON = False to enter manual

zero/span mode. Verify

mode_status has changed from Normal EC mode to

Manual Zero/Span mode.

If the CPEC200 program is not configured for valve operation,

mode_status will report Standby mode (Pump is Off). See

Appendix A, CPEC200 Diagnostics, for further details.

Tip: Normally a manual zero/span is started when the CPEC200 is in the EC

mode (pump on). In that case, turn the pump off to begin a manual zero/span.

However, if the zero/span sequence is running, stop the sequence by turning

the pump on. This will return the system to normal EC mode. Then turn the

pump off to enter manual zero/span mode.

Set

DO_P_offset = True. This will turn all of the valves off and command the

CPEC200 to measure the pressure offset. Watch the value of

press_offset.

Turning the pump on or setting

DO_P_offset = True will reset this variable to

zero. Within approximately 10 s after turning the pump off and setting

DO_P_offset = True), press_offset will be set to the value of cell_press (with

a 5 s moving average) and

DO_P_offset will reset to False. Wait until

DO_P_offset = False before proceeding.

Tip: The purpose of this step is for the CPEC200 to store the ambient pressure

(measured by

cell_press when there is no flow in the system) in public variable

press_offset. This allows the CPEC200 to infer the flow in the sample cell

from the pressure rise resulting from the zero/span flow. Because ambient

pressure changes over time (and the pressure sensor may drift over time), this

pressure must be measured each time a zero/span is performed.

6.3.2.2 Checking and Setting the Zero

The keyboard display has menus to facilitate manual zero control for users that

are onsite. Navigate:

Manual zero/span → Manual Control → H2O and CO2 Zero

NOTE

37