Appendix H. CPEC200 Pump

Replacement

H.1 Introduction

A properly maintained CPEC200 system will exceed the lifetime of the

system’s pump. This section provides step-by-step instructions for the user to

replace the system pump (pn 26402), rather than needing to return the pump

enclosure to Campbell for replacement.

H.2 Removal

To remove a CPEC200 pump, carry out the following steps:

1. Place the pump module on a horizontal surface.

It is very difficult to replace the pump if the module is mounted

vertically on a tower.

2. Remove the four, #10 self-sealing screws that hold the filter assembly

within the CPEC200 pump module enclosure (see FIGURE H-1). If

these screws become lost or damaged, replace them with pn 26412.

FIGURE H-1. Four screws holding filter assembly inside CPEC200

pump module enclosure

3. Stand the filter up in the enclosure and remove the tubing from the

filter inlet by loosening the nut with an 11/16” wrench. Remove the

tubing from the outlet of the filter by loosening the nut with a 9/16”

wrench (FIGURE H-2).

The filter cartridge in the pump module is unlikely to clog over

the lifetime of the CPEC200 system. If the filter holder or

cartridge become damaged, however, the replacement parts are pn

26407 (filter holder) and pn 26408 (filter cartridge).

NOTE

NOTE