beautypg.com

1 enclosure desiccant, 2 ec155 intake filter, 3 ec155 windows – Campbell Scientific CPEC200 Closed-Path Eddy-Covariance System User Manual

Page 50: Enclosure desiccant, Ec155 intake filter, Ec155 windows

background image

CPEC200 Closed-Path Eddy-Covariance System

7.1 Enclosure Desiccant

Check the humidity indicator card in the mesh pocket in the CPEC200 system

enclosure door and the EC100 enclosure door. The humidity indicator card has

three colored circles that indicate the percentage of humidity (see FIGURE

4-14). Desiccant packets inside the enclosure should be replaced with fresh

packets when the upper dot on the indicator begins to turn pink. The indicator

card does not need to be replaced unless the colored circles overrun.

Campbell Scientific strongly suggests replacing desiccant

instead of reactivating old desiccant. Improper reactivation

can cause the desiccant packets to explode.

The replacement desiccant pack is the 4905 Single 4-Unit Silica Desiccant Bag

which can be purchased in quantities of 20 as pn 6714. The replacement

humidity indicator card is pn 28878. See Section 4.1.5, Replacement Parts, for

more detail.

If the desiccant packs in the CPEC200 system enclosure are requiring frequent

replacement, check that the feedthrough cap is properly installed. In very

humid conditions it may be helpful to seal the cable feedthrough with

plumber’s putty as described in Section 5.3.4, Apply Power.

7.2 EC155 Intake Filter

Over time, the EC155’s intake filter will become plugged with particulates

from the air causing an increased pressure drop. The EC155 has a differential

pressure sensor with a ±7 kPa range to measure the sample cell pressure with

respect to ambient pressure. As the intake filter plugs over time, the

differential pressure will decrease from approximately –3 kPa (typical for a

clean filter at 7 LPM). If the differential pressure reaches –7 kPa, the data will

no longer be valid. It is important to monitor the differential pressure and

replace the filter before this limit is reached.

The default replacement filter is pn 26072. It is a 2.5-cm (1.0-in) diameter,

sintered stainless steel disk filter of 20 µm pore size, with a molded

Santoprene™ shell. An alternative 40 µm filter (pn 28698) is also available.

Choose the 40 µm filter if the default 20 µm filter clogs long

before the EC155 optical windows become dirty.

See the EC155 CO

2

and H

2

O Closed-Path Gas Analyzer Manual for details on

replacing the intake filter.

7.3 EC155 Windows

Over time, the optical windows of the EC155 will become dirty and will need

to be cleaned. To evaluate the cleanliness of the windows, monitor the CO

2

signal and the H

2

O signal. These will have a value of approximately 1.0 for

clean windows, but will decrease as the windows become dirty. The EC155 is

calibrated for a range of signal levels down to 0.80. Clean the windows as

instructed in the EC155 CO

2

and H

2

O Closed-Path Gas Analyzer Manual

before the CO

2

and H

2

O signals reach 0.80.

CAUTION

NOTE

40