beautypg.com

F.3 c codes, C codes, Appendix f - error codes – MagTek DynaPro Mini Programmer's99875629 User Manual

Page 97

background image

Appendix F - Error Codes

DynaPro Mini| PIN Encryption Device | Programmer’s Reference (Commands)


Page 97

Code

Calculated

Sn where n

= sum of

the

following

Description

Meaning

+2

Keypad calibration not complete
(Status stored in BPK)

After reset, calibration should be
redone (but really only status has been
lost)

+4

MSR key pairing not completed

Need to perform action.
Permanent.

+8

Tamper sensors not activated

Status set after BPK initialized.
Cleared once sensors activated.
Stored in BPK

+16

Keypad activation sequence not sent

Need to perform action.
Permanent.

+32

Keypad tamper detected during power up
(Threshold stored in BPK)

Retry. Keypad may have shifted or
been modified. Something nearby may
be interfering. Keypad calibration bad
or not performed yet.

Table 3-91 - Common "S" Error Codes

Common

S codes

Description

Meaning

S63/S31

Usually new unit

Usually occurs after battery reset or battery
power loss

S43/S35

Keypad calibration not complete
(Status stored in BPK)

After reset, calibration should be redone
(but really only status has been lost)

F.3 C Codes

Table 3-92 - "C" Error Codes

Code Calculated

Cn where n = sum of the

following

Description

+1

Mfg Unbind certificate not loaded

+2

Device CA certificate not loaded

+4

PIN CA certificate not loaded

+8