7 . c – DAVIS Weather Monitor II (7440, 7440CS) User Manual
Page 40

Calibration Numbers
Page 36
Weather Monitor II
7 . C
A L I B R A T I O N
N
U M B E R S
Calibration (CAL) numbers are used to turn raw data from the sensors into 
meaningful readings on the display. The standard or default CAL numbers are 
automatically entered by the unit when it is first powered up and whenever 
power is lost and later restored.
Calibration numbers should not be changed except for certain specialized 
applications. This section explains why you may wish to change CAL, the 
equations used by the Weather Monitor II, and the default values for CAL 
numbers. For instructions on changing CAL, see “Changing Calibration Num-
bers” on page 37.
Note:
We do not recommend changing the preset calibration numbers unless you are using a metric rain 
collector.
I
NSIDE
AND
O
UTSIDE
T
EMPERATURE
Change CAL if you wish to adjust the Weather Monitor II so that is reads the 
same as another thermometer. Note that there are individual CAL numbers for 
both inside and outside temperature. If you change CAL for outside tempera-
ture, the wind chill reading will also be affected.
C
ALIBRATED
T
EMPERATURE
= T
EMPERATURE
+ CAL
Default CAL = 0 (for ˚F); 0 (for ˚C).
W
IND
S
PEED
The four units of measure provided are generally sufficient for most applica-
tions. If you change CAL for wind speed, the wind chill readings will be incor-
rect.
C
ALIBRATED
W
IND
S
PEED
= R
EVOLUTION
X
(1600 / CAL)
Default CAL = 1600 (for mph); 995 (for kph); 1843 (for knots); 3578 (for m/s).
H
UMIDITY
Change CAL if your outside humidity reading is off by a consistent amount. A 
CAL number exists for outside humidity only. If you change CAL, dew point 
readings will also be affected.
C
ALIBRATED
H
UMIDITY
= H
UMIDITY
+ CAL
Default CAL = 0%.