5.2 Switching off the HygroGen2
There are two Shutdown procedures; these are presented after you press the on-screen
power button. It is important that these are observed; instrument faults attributable to
incorrect Shutdown are not covered by the warranty.
Shut Down
In normal circumstances use “Shutdown”. This dries the chamber to prevent residual water and
isolates the humidifier from the chamber. Once correctly shutdown, the HygroGen2 is in a suitable
condition for short- to medium-term periods without use. It is also put into a condition where it can be
moved carefully in a horizontal position (e.g. on a trolley) to a different location (say, within a
building).
Shutdown Procedure:
1. Press the “Power Off” button on-screen.
2. Select “Shut Down”.
3. Allow the HygroGen2 to shut down. The HygroGen2 will isolate the humidifier
and cool/dry to ensure excess water is removed. This may take several minutes.
DO NOT switch off at the wall before the HygroGen2 has shut down.
DO NOT switch off at the back of the HygroGen2 before it has shut down.
Figure 34: Shutdown Screen