4.2.3 Corrections
When getting your dew point hygrometer calibrated, please request a dew point calibration with
corrections from your calibration supplier and a table of corrections for applied set-points for your
external temperature probe. Enter the correction values from the instrument’s calibration certificate
via the two tables found via the “Corrections” option in the “External” drop-down menu and press the
toggle button so that it displays “Applied”.
Figure 26: External Reference Corrections
4.2.4 Calibration Info
Information about the calibration of your external reference can be entered via the “Calibration Info”
option in the “External” drop-down menu.
Figure 27: Reference Calibration Information