Section 4.2, E-m-hg2-s-v2.1 – ROTRONIC Hygrogen2 User Manual
Page 35

E-M-HG2-S-V2.1
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
HygroGen2: Humidity and Temperature Generator
with AutoCal/AutoCal+, Remote Control and
Range Extensions options.
Instruction Manual
Instruction Manual for Software
Version 2.1
Document Type
Page 31 of 75
Document title
Figure 20: Edit AutoCal Reference Information
Reference calibration information can be entered here, including the last and next-due calibration dates
and any comments.
When an AutoCal run is started, the user is warned should the reference calibration be out of date and
prompted to update the reference information should the reference probe have been swapped out. The
user is further asked to confirm that the operator name to be printed on the certificate does not require
changing. If a probe that has previously been used and has reference information stored on the
HygroGen2 is restored to the instrument, the information will automatically be reapplied.
© 2014; Rotronic AG
E-M-HG2-S-V2.1