3.4.4 Edit Program Screen
1) Graphic Indicator
2) Ramp/Soak control
3) Temperature set-point
4) Humidity set-point
5) Step duration
6) Insert Step
7) Delete Step
8) Scroll to view Steps
9) Program title
10) Temperature tolerance
11) Humidity tolerance
12) Temperature Only
13) Programmer
Select a Program to Edit
Select the desired program by clicking on the program button and entering the desired program
number.
Edit a Program Name
To edit the program name click on it (USB keyboard required) (item 9).
Edit a Step
Each program is made of steps. Each step can be edited. Enter the desired step set-points and duration
(items 3, 4 & 5). A graphical representation of the step is shown to help visualise the sequence.
Insert or Delete a Step (items 6 & 7)
Use the insert and delete buttons to add and remove steps. If there are more than 7 steps in the program
use the scroll bar (item 8) to view.
Ramp or Soak (item 2)
Set temperature and humidity to either ramp or soak using item 2. A soak is where the HygroGen2
maintains the temperature and humidity in the chamber constant for the duration of the step, within
the tolerances set (items 10 & 11). A ramp is where the HygroGen2 changes the selected set point at a
rate determined by the subsequent set-point value and the current step duration.
Figure 11: Edit Program