3.2 Upgrading From Previous Versions
It is strongly recommended to upgrade your HygroGen2 to at least software version 2.0. Version 2.x
contains numerous core system improvements over version 1.x.x, including robustness against power
outages, water level sensor calibration and compatibility improvements. Each version is a free upgrade.
If you are upgrading the HygroGen2 from 1.x.x, it is strongly recommended the HygroGen2 is serviced.
This will allow hardware updates to maximise the performance of the V2.x.x software. These include
connecting the internal network cable and updating the power shutdown hardware.
The Enhanced Features described in
are unlocked by purchasing the appropriate software
code from your ROTRONIC dealer.
To install version 2.1, you will first need to install version 2.0.
3.2.1 Installation of HygroGen2 version 2.0 from recovery installation USB stick
Version 2.0 recovery installation USB sticks (HG2-V2.0.UPGRADE) can be obtained from your
ROTRONIC Dealer.
A recovery installation USB stick can also be made using a blank 4GB USB stick and
Step by step instructions for preparing the USB stick are included in the zipped file. When you have
downloaded the file and prepared the USB stick, follow the instructions below.
You will need a USB keyboard and mouse plugged into the HygroGen2 to perform the following steps.
Please make sure your HygroGen2 is connected to a stable power supply.
i.
Make a note of existing HW4 software version and serial number.
a. Boot your HygroGen2 in the conventional manner. Press the HW4 button and “Start HW4”.
b. Select “About HW4” from the “Help” menu.
c. Make a note of the HW4 version number and serial number (ID) which should be in the
format XX YYYYYYYY ZZZZZ.
d. Close HW4.
e. Insert the USB stick into one of the back ports of the HygroGen2.
f. Go to the settings screen and press Update, then press OK to restart your HygroGen2.
ii.
HygroGen2 version 2.0 installation
The HygroGen2 will now boot with a Windows Loading screen and load with a HYGROGEN
INSTALLATION DRIVE splash screen. It will take a few moments to initialise all the hardware
components.