15 TROUBLESHOOTING & ERROR MESSAGES


TROUBLESHOOTING

Problem Possible

Cause

Solution

Power ON but no display

AC adapter not connected properly

Connect AC adapter properly

Unstable Reading

Electrode submersion in sample is
insufficient

Place electrode deeper into sample

Broken

electrode

Replace electrode

External noise causing instability in
reading.

Remove or switch off interfering motor

Dirty electrode

Clean Electrode. Reactivate if necessary

Slow response

Dirty electrode

Clean Electrode. Reactivate if necessary

ERROR MESSAGES

LCD Display

Indicates

Cause

Solution

Error:
EEPROM Read Error

Memory write error

Unit is too old (> 10 years)

Turn meter off and return to authorized
distributor if necessary

Error:
EEPROM write Error

Memory checksum
error

Hardware failure

Turn meter off and return to authorized
distributor if necessary

Error:
DTR is set too high

DTR line is
disconnected

Hardware fault

Turn meter off and return to authorized
distributor if necessary

Error:
Communication error

Frame/parity errors

Turn meter off and return to authorized
distributor if necessary

Error:
Buffer Overflow

Received buffer
overflowing

Turn meter off and return to authorized
distributor if necessary

Error:
pH Offset Error

pH offset too much

pH offset should not exceed
±1 pH from the display

Check electrode condition

Make sure buffer is fresh and correct

Recalibrate meter

Error:
pH Slope Error

Incorrect pH slope

pH slope is out of allowable
limit

Check electrode condition

Make sure buffer is fresh and correct

Recalibrate meter

Error:
Ion Slope Error

(pH 2100 meter only)

Incorrect ion slope

Average ion slope is not
within 10mV-75mV/decade

Check electrode condition

Make sure buffer is fresh and correct

Recalibrate meter

Error:
mV Offset too large

Offset too much

Value input is not within
±150.0mV from the display

Recalibrate meter

Error:
Out of Range

Out of Range

Value input exceeds the
specified range

Re-enter value

Error:
ATC offset too much

Offset of temp. probe
too much

Temp. calibration offset is
not within ±5

o

C (or ±9

o

F)

Re-enter value

Error:
Too much offset@7pH

Offset too much

pH offset should not exceed
±1 pH from the display
(custom pH buffer)

Re-enter value

Error:
Slope out:75%-105%

Incorrect pH slope

pH slope is not within 75% to
105% (custom pH buffer)

Re-enter value

Error: Pts too close
Or slope diff. sign

Incorrect slope
alignment

Slope is not negative or cal.
Points are not ±1 pH apart
(custom pH buffer)

Re-enter value