15 TROUBLESHOOTING & ERROR MESSAGES
TROUBLESHOOTING
Problem Possible
Cause
Solution
Power ON but no display
AC adapter not connected properly
Connect AC adapter properly
Unstable Reading
Electrode submersion in sample is
insufficient
Place electrode deeper into sample
Broken
electrode
Replace electrode
External noise causing instability in
reading.
Remove or switch off interfering motor
Dirty electrode
Clean Electrode. Reactivate if necessary
Slow response
Dirty electrode
Clean Electrode. Reactivate if necessary
ERROR MESSAGES
LCD Display
Indicates
Cause
Solution
Error:
EEPROM Read Error
Memory write error
Unit is too old (> 10 years)
Turn meter off and return to authorized
distributor if necessary
Error:
EEPROM write Error
Memory checksum
error
Hardware failure
Turn meter off and return to authorized
distributor if necessary
Error:
DTR is set too high
DTR line is
disconnected
Hardware fault
Turn meter off and return to authorized
distributor if necessary
Error:
Communication error
Frame/parity errors
Turn meter off and return to authorized
distributor if necessary
Error:
Buffer Overflow
Received buffer
overflowing
Turn meter off and return to authorized
distributor if necessary
Error:
pH Offset Error
pH offset too much
pH offset should not exceed
±1 pH from the display
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Error:
pH Slope Error
Incorrect pH slope
pH slope is out of allowable
limit
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Error:
Ion Slope Error
(pH 2100 meter only)
Incorrect ion slope
Average ion slope is not
within 10mV-75mV/decade
Check electrode condition
Make sure buffer is fresh and correct
Recalibrate meter
Error:
mV Offset too large
Offset too much
Value input is not within
±150.0mV from the display
Recalibrate meter
Error:
Out of Range
Out of Range
Value input exceeds the
specified range
Re-enter value
Error:
ATC offset too much
Offset of temp. probe
too much
Temp. calibration offset is
not within ±5
o
C (or ±9
o
F)
Re-enter value
Error:
Too much offset@7pH
Offset too much
pH offset should not exceed
±1 pH from the display
(custom pH buffer)
Re-enter value
Error:
Slope out:75%-105%
Incorrect pH slope
pH slope is not within 75% to
105% (custom pH buffer)
Re-enter value
Error: Pts too close
Or slope diff. sign
Incorrect slope
alignment
Slope is not negative or cal.
Points are not ±1 pH apart
(custom pH buffer)
Re-enter value