10.2.5

pH Measurement Alarm Setup Program P1.3


Use this program to select alarm for low and high pH
measurements.

1. Enter

pH/Temperature

setup

mode as described in steps 1-4

in 10.2.1 Entering pH/temperature setup mode (page 30).


2. Use

/MI to scroll to program pH Meas. Alarm: P1.3. See

Figure 87.


3. Press

ENTER to enter program pH Meas. Alarm: P1.3.


4. Press

>/CAL to toggle between pH Alarm On and Off. The

selection will flash. See Figure 88.


5. Press

ENTER to confirm selection.

If you selected On continue to step 6

• If you select Off, pressing ENTER returns to the setup

mode.

Press

/MI key to continue to the next setup program, or press

/MR to go to the previous program. Press EXIT to return to
the pH measurement mode.


6. If

you

selected

On, the display now reads Key in Low Alarm

pH Set Point. Key in the low pH alarm set point using the
numeric keypad. If you make a mistake, use the

/MI or ∇/MR

keys to highlight mistake, then re-enter. See Figure 89. Press
ENTER to confirm keyed in value.


7. The display now reads Key in High Alarm pH Set Point. Key in

the high pH alarm set point using the numeric keypad. The
high alarm must be set at least 1.0 pH greater than the low
alarm. If you make a mistake, use the

/MI or ∇/MR keys to

highlight mistake, then re-enter. See Figure 90.


8. Press

ENTER to confirm high pH set point value and return to

the setup mode.

Press

/MI key to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to the pH measurement mode.

Figure 89

Figure 90

Key In Low Alarm

pH Set Point:2 . 000

Key In High Alarm

pH Set Point:14 . 000

pH Meas. Alarm :P1.3
Press ENT. to Select

Figure 87

Figure 88

pH Alarm On
Select & Press ENTER

Off