10.5.4
Data log setup mode P4.2
Use this program select automatic or manual data logging.
Automatic data logging can be set at timed intervals from 10
seconds to 23 hours, 59 minutes, 59 seconds.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use
∆/MI key to scroll to program Data Log Setup: P4.2.
See Figure 170.
3. Press
ENTER to enter program Data Log Setup: P4.2.
4. Press
>/CAL key to toggle between Data Log Auto and
Manual. The selection will flash. See Figure 171.
5. Press
ENTER to confirm selection.
•
If you selected Auto, continue to step 6.
• If you select Manual, pressing ENTER returns to the setup
mode. Press
∆/MI key to continue to the next setup
program, or press
∇/MR to go to the previous program.
Press EXIT to return to measurement mode.
6. If you selected Auto in step 5, the display now reads Auto
DataLog Period Hr: Min: Sec : Key in the time period
(minimum of 10 seconds to maximum of 23 hours, 59 minutes,
59 seconds) using the numeric keypad. If you make a mistake,
use
∆/MI or ∇/MR keys to highlight mistake, then re-enter.
See Figure 172.
7. Press
ENTER to confirm selection. The screen flashes Sys
Updating... Please wait... and then returns to the setup mode.
Press
∆/MI key to continue to the next setup program, or
press
∇/MR to go to the previous program. Press EXIT to
return to measurement mode.
10.5.5
Memory recall setup mode P4.3
Use this program to select automatic or manual memory recall. See
page 24 for more information on memory recall.
1. Enter meter setup mode as described in steps 1-4 in Section
10.5.1 Entering meter setup mode (page 50).
2. Use
∆/MI key to scroll to program MemRecall Setup: P4.3.
See Figure 173.
3. Press
ENTER to enter program MemRecall Setup: P4.3.
4. Press
>/CAL key to toggle between Memory Recall Mode
Auto and Manual. The selection will flash. See Figure 174.
5. Press
ENTER to confirm selection and return to the setup
mode. Press
∆/MI key to continue to the next setup program,
or press
∇/MR to go to the previous program. Press EXIT to
return to measurement mode.
Auto DataLog Period
Hr:1 Min:30 Sec:0
Data Log:Auto Manual
Select & Press ENTER
Data Log Setup :P4.2
Press ENT. to Select
Figure 170
Figure 171
Figure 172
Memory Recall Mode
Auto Manual
MemRecall Setup:P4.3
Press ENT. to Select
Figure 173
Figure 174