10.4.6

Ion mode setup program P3.4


Use this program to select high or low ion resolution mode. High
resolution has ±500 mV value, low has ±1850 mV value.

1. Enter ion setup mode as described in steps 1-4 in Section

10.4.1 Entering ion setup mode (page 43).


2. Use

/MI key to scroll to program Ion Mode Setup: P3.4. See

Figure 144.


3. Press

ENTER to enter program Ion Mode Setup: P3.4.


4. Press

>/CAL to toggle between High or Low. The unit

selected will flash. See Figure 145.


5. Press

ENTER to confirm selection and return to the setup

mode.


6. Press

/MI key to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to the ion measurement mode.

Resolution: High Low

Select & Press ENTER

Ion Mode Setup : P3.4

Press ENT. to Select

Figure 144

Figure 145