Instruction Manual

pH 1100/ 2100

23

6 MEMORY

FUNCTIONS

6.1 Memory

input


This meter can store up to 100 sets of data in several combinations:


pH 1100 meter:
• pH, temperature, date and time
• mV, temperature, date and time

pH 2100 meter:

• pH, temperature, date and time
• mV, temperature, date and time
• ion, mV, date and time


For example, you can store 32 pH and 68 mV values for the pH 1100 or 50 ion, 25 mV and 25 pH values for the pH 2100.
This meter can be setup to log readings automatically or manually. To set the meter to log automatically, see Meter
Setup Program P4.2, Data Log Setup
(page 52). The meter can log readings at intervals from 10 seconds to 23 hours 59
minutes and 59 seconds.

6.1.1

Manually storing a reading into memory


1. During any measurement mode function, press

/MI key to

input data into memory.


2. The upper display will read current measurement function and

the lower display will flash Data Log.... Loc: (the memory
storage location). See Figure 61. The meter then returns to
measurement mode.


NOTE:
If the memory is full, the first value stored will be erased to create space for the new value.

Figure 61

5.695pH 25.3 C ATC
Data Log. . . . Loc: 9

°