beautypg.com

Ion conc: ppm abs. mv : 100.7mv, Ion conc:0.01ppm 1 point cal done, Ion conc:0.10ppm 2 point cal done – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual

Page 20

background image

Instruction Manual

pH 1100/ 2100

20

5

ION CALIBRATION AND MEASUREMENT (ONLY APPLICABLE FOR pH 2100)

The pH 2100 meter reads ion concentration in ppm, ppt, mg/l, g/l, mmol/l and mol/l units. See Ion setup Program P3.0
(page 43) to select ion concentration units.

5.1 Ion

calibration

This instrument is capable of up to 5-point custom ion calibration with a minimum of 2-point calibration. Custom
calibration values can be any value between 0.001 and 19998. Once the first calibration value is keyed in, all other
calibration will automatically be at least one decade apart.

For example if your first calibration point is at 0.001, all other calibration points will be at least one decade apart from
0.001 (0.001, 0.01, 0.1,1.0, 10.0, 100, 1000, 10000).

For best accuracy calibrate your meter to points with similar concentrations to the solutions you want to test.

1. Connect an Ion Selective Electrode (ISE) to the BNC input

connector on the back of the meter.


2. Turn

the

meter

ON. If the meter is not in the Ion measurement

mode, press EXIT and use >/CAL to select Ion. The mode
selected will flash. See Figure 48. Press ENTER to confirm
selection. The meter is now in Ion measurement mode.


3. If the meter has not been calibrated or has been reset, the

display will flash Ion Meas and current mV value. See Figure
49
. If the meter has been calibrated previously, the display will
show the current concentration value plus the absolute mV
produced by the electrode.


4. Dip the electrode into the first calibration standard. Start with

the calibration standard that has the lowest concentration and
move up to higher concentration standards.


5. Press

>/CAL to enter the ion calibration mode. The display will

flash Cal Mode Activated, Please wait... See Figure 50. The
display then shows Ion Conc: ppm (or unit selected in ion
setup mode) in the upper display and Abs. mV: in the lower
display. See Figure 51.


6. The first calibration point must be keyed in by the user. Key in

the concentration value of the standard using the numeric
keypad. If you make a mistake, use the

/MI or ∇/MR keys to

highlight mistake, then re-enter.


7. Press

ENTER to confirm the first calibration point. The display

will flash 1 Cal Point Done *, then the meter proceeds to the
next calibration point at least 1 decade apart from the first
calibration point. See Figure 52.


8. Use

/MI or ∇/MR keys to move the cursor to next calibration

point. All points will be a decade apart from the first calibration
point.


9. Rinse off the electrode with deionized water before placing it

in the next calibration standard.


10. Allow probe to stabilize in calibration standard. Press ENTER

to confirm. The display flashes 2 Point Cal Done *. See
Figure 53.
The meter proceeds to the next calibration point.

pH/Temp mV Ion

Select & Press ENTER

Cal Mode Activated
Please wait. . .

Ion Meas 160.3mV
Jan 20 2002 21:54:34

Figure 48

Figure 49

Figure 50

Figure 51

Figure 52

Figure 53

Ion Conc: ppm
Abs. mV : 100.7mV

Ion Conc: ppm
Abs. mV : 100.7mV

Ion Conc:0.01ppm

1 Point Cal Done

*

Ion Conc:0.10ppm

2 Point Cal Done

*