Cal alarm due: on off select & press enter, Ioncaldue alarm :p3.2 press ent. to select – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual

Page 45

background image

Instruction Manual

pH 1100/ 2100

45

10.4.4

Ion calibration due alarm setup program P3.2


Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.

1. Enter ion setup mode as described in steps 1-4 in Section 10.4.1

Entering ion setup mode (page 43).


2. Use

/MI key to scroll to program IonCalDue Alarm: P3.2.

See

Figure 138.


3. Press

ENTER to enter program IonCalDue Alarm: P3.2.


4. Press

>/CAL key to toggle between Cal Alarm Due: On and Off.

The selection will flash. See Figure 139.


5. Press

ENTER to confirm selection.

If you selected On, continue to step 6.

If you select Off, pressing ENTER returns to the setup mode.


Press

/MI key to continue to the next setup program, or press

/MR to go to the previous program. Press EXIT to return to mV
measurement mode.


6. If you selected On in step 5, the display reads Key in Cal Due

Period (1-365 days) = 1. Key number of days using the numeric
keypad. If you make a mistake, use

/MI or ∇/MR keys to

highlight mistake, then re-enter. See Figure 140.


7. Press

ENTER to confirm number of days to next calibration;

screen flashes Sys Updating, Please wait… and returns to the
setup mode. Press

/MI key to continue to the next setup

program, or press

/MR to go to the previous program. Press

EXIT to return to mV measurement mode.

Keyin Cal Due Period

(1-365 days) = 1

Cal Alarm Due: On Off

Select & Press ENTER

IonCalDue Alarm :P3.2
Press ENT. to Select

Figure 138

Figure 139

Figure 140