10.2.7

Stability Setup Program P1.5


Use this program to turn on or off the pH stability indicator and to
set level of stability (low, medium or high).

Stability definitions

High stability:
reading is stabilized quickly, but reading is less accurate.
Medium stability:
reading stability is averaged between high and low stability.
Low stability:
reading is stabilized slowly, but guarantees high accuracy

1. Enter pH/Temperature setup mode as described in steps 1-

4 in 10.2.1 Entering pH/temperature setup mode (page 30).


2. Use

/MI to scroll to program Stability Setup: P1.5.

See

Figure 94.


3. Press

ENTER to enter program Stability Setup: P1.5.


4. Press

>/CAL key to toggle between pH Stable Criteria: On

and Off. The selection will flash. See Figure 95.


5. Press

ENTER to confirm selection.

If you selected On, continue to step 6.

• If you select Off, pressing ENTER returns to the setup

mode.

Press

/MI to continue to the next setup program, or press

/MR to go to the previous program. Press EXIT to return to
the pH measurement mode.


6. If you selected On in step 5, the display now reads Degree

of Stability: Low Medium High. Press >/CAL to toggle
between Low, Medium and High. The selection will flash.
See Figure 96.


7. Press

ENTER to confirm selection and return to the setup

mode. Press

/MI to continue to the next setup program, or

press

/MR to go to the previous program. Press EXIT to

return to the pH measurement mode.

Mode

Degree of Stability
Low Medium High

Stability Setup : P1.5
Press ENT. to Select

pH Stable Criteria
On Off

Figure 94

Figure 95

Figure 96