10.3.6
mV calibration data reset setup program P2.4
Use this program to erase all mV calibration data.
1. Enter mV setup mode as described in steps 1-4 in Section
10.3.1 Entering mV setup mode (page 37).
2. Use
∆/MI key to scroll to program mV Cal Reset: P2.4.
See Figure 125.
3. Press
ENTER to enter program mV Cal Reset: P2.4.
4. Press
>/CAL key to toggle between mV Reset: Yes and No.
The selection will flash. See Figure 126.
5. Press
ENTER to confirm selection.
• If you selected Yes, the screen flashes mV being Reset,
Please wait... See Figure 127. The meter then returns to
setup mode.
• If you select No, the screen flashes Reset Aborted and
then returns to the setup mode. See Figure 128. This is
the last mV setup program. Press
∇/MR to go to the
previous program. Press
∆/MI to go back to the first mV
setup program. Press EXIT to return to mV measurement
mode.
Reset Aborted
mV being Reset
Please wait. . .
mV Reset: Yes No
Select & Press ENTER
mV Cal Reset :P2.4
Press ENT. to Select