10.3.6

mV calibration data reset setup program P2.4


Use this program to erase all mV calibration data.


1. Enter mV setup mode as described in steps 1-4 in Section

10.3.1 Entering mV setup mode (page 37).


2. Use

/MI key to scroll to program mV Cal Reset: P2.4.

See Figure 125.


3. Press

ENTER to enter program mV Cal Reset: P2.4.


4. Press

>/CAL key to toggle between mV Reset: Yes and No.

The selection will flash. See Figure 126.


5. Press

ENTER to confirm selection.

• If you selected Yes, the screen flashes mV being Reset,

Please wait... See Figure 127. The meter then returns to
setup mode.

• If you select No, the screen flashes Reset Aborted and

then returns to the setup mode. See Figure 128. This is
the last mV setup program. Press

/MR to go to the

previous program. Press

/MI to go back to the first mV

setup program. Press EXIT to return to mV measurement
mode.




Reset Aborted

mV being Reset
Please wait. . .

mV Reset: Yes No
Select & Press ENTER

mV Cal Reset :P2.4
Press ENT. to Select