Cal alarm due: on select & press enter, Off phcaldue alarm : p1.4 press ent. to select – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual

Page 33

background image

Instruction Manual

pH 1100/ 2100

33

10.2.6

pH Calibration Due Alarm Setup program P1.4


Use this program to set a reminder when the next calibration is due.
Choose from 1 to 365 days from last calibration.

1. Enter pH/Temperature setup mode as described in steps 1-4 in

10.2.1 Entering pH/temperature setup mode (page 30).


2. Use

/MI to scroll to program pHCalDue Alarm: P1.4. See

Figure 91.


3. Press

ENTER to enter program pHCalDue Alarm: P1.4.


4. Press

>/CAL to toggle between Cal Alarm Due:On and Off. The

selection will flash. See Figure 92.


5. Press

ENTER to confirm selection.


If you selected On, continue to step 6

If you select Off, pressing ENTER returns to the setup mode.


Press

/MI to continue to the next setup program, or press ∇/MR

to go to the previous program. Press EXIT to return to the pH
measurement mode.


6. If you selected On in step 5, the display reads Key in Cal Due

Period (1-365 days) =. Key number of days using the numeric
keypad. If you make a mistake, use the

/MI or ∇/MR keys to

highlight mistake, then re-enter. See Figure 93.

7.

Press ENTER to confirm number of days to next calibration,
screen flashes Sys Updating, Please wait... and returns to the
setup mode. Press

/MI key to continue to the next setup

program, or press

/MR to go to the previous program. Press

EXIT to return to the pH measurement mode.

Keyin Cal Due Period
(1-365 days) = 1

Figure 91

Figure 92

Figure 93

Cal Alarm Due: On

Select & Press ENTER

Off

pHCalDue Alarm : P1.4
Press ENT.

to Select