Setup mode activated please wait, Usa nst bf1 bf2 cust select & press enter – Thermo Fisher Scientific CyberScan pH 1100/pH 2100 User Manual
Page 30
Instruction Manual
pH 1100/ 2100
30
10.2 pH/ temperature setup mode
10.2.1
Entering pH/temperature setup mode
1. Turn the meter ON. If the meter is not in the pH measurement
mode, press EXIT and use >/CAL to select pH/Temp. The
mode selected will flash. Press ENTER to confirm. The meter is
now in pH measurement mode.
2. Press
SETUP to enter the setup mode. The display will flash
Setup Mode Activated, Please wait... See Figure 79.
3. The meter then enters the pH/Temp Meter Setup screen. Use
>/CAL to select pH/Temp. The mode selected will flash. See
Figure 80. Press ENTER to enter pH/Temp setup mode.
4. The first setup program in pH/Temp mode is pH Buffer Setup:
P1.0. See Figure 81. If you want to skip to another program,
use the
∆/MI and ∇/MR keys to scroll to the desired program.
5. To exit the setup mode at anytime, press EXIT. The meter
returns to the pH/Temp measurement mode.
NOTE: To exit the setup mode at anytime without confirming
changes, press EXIT. If you have made any changes to any
programs and press EXIT, these changes will be confirmed. To
reset meter to default factory settings, see setup program Meter
Reset P4.7 (page 54).
10.2.2
pH buffer setup program P1.0
Use this program to select calibration buffer sets: USA, NIST, Bf1, Bf2 or custom.
Selectable buffer sets:
USA: 1.68, 4.01, 7.00, 10.01, 12.45
NIST: 1.68, 4.01, 6.86, 9.18, 12.45
Bf1: 1.68, 4.01, 7.01, 9.18, 12.45
Bf2: 1.68, 4.01, 6.86, 10.01, 12.45
Custom: any 5 user selected custom buffers
1. Enter the pH/Temperature setup mode as described above
(10.2.1 Entering pH/temperature setup mode). The first setup
mode is pH Buffer Setup: P1.0. See Figure 81.
2. Press
ENTER to enter program pH Buffer Setup: P1.0.
3. Press
>/CAL key to toggle between USA NST Bf1 Bf2 Cust.
Buffer set selected will flash. See Figure 82.
4. Press
ENTER to confirm selection and return to the setup
mode. Press
∆/MI key to continue to the next setup program,
or press
∇/MR to go to the previous program (P1.7 pH
Calibration Reset). Press EXIT to return to the pH
measurement mode.
Setup: pH/Temp Meter
Select & Press ENTER
pH Buffer Setup:P1.0
Press ENT. to Select
Setup Mode Activated
Please wait. . .
Figure 79
Figure 80
Figure 81
Figure 82
USA NST Bf1 Bf2 Cust
Select & Press ENTER